Send Message
Home News

FAQ for IP Water Ingress Protection Testing Equipment

China Pego Electronics (Yi Chun) Company Limited certification
China Pego Electronics (Yi Chun) Company Limited certification
Je viens de recevoir les doigts d essais, Je suis très satisfait , c est très bien

—— ESTI Formation

Dear Alice, The machines are working well now. Thank you for your support and patience to solve our problems and give us very useful suggestions in negotiation. Really a good cooperation.

—— TUV Rheinland Pvt. Ltd

Dear Penny, We really appreciate to buying from Pego Group for several years, all products come with good quality amd easy operation.

—— LIA Laboratories Limited

I have received the spring hammers, they are great!

—— BSH Group

Yes we got the probes and our engineers like them. Thanks.

—— Gamma Illumination Pty Ltd

the spring hammer looks good and works fine

—— Sigma Connectivity

This is to confirm the order has been well received and all is in good order. May I thank you for your very professional service.This has been a pleasant transaction.

—— Gallagher Group

절연강도 시험기 잘 받았습니다. 빠른 배송에 감사드리며 차후 또 다른 제품구입시 연락드리겠습니다.

—— 유진코퍼레이션에 황동익입니다.

I am sending this email to let you know that we receive the 2 vessels with the calibration certificate. I would like to thank you for the quality of the vessels and for your good work. I can’t wait to work with you on other topics.

—— Brandt France

we can confirm that it works very well! thank you for the good quality equipment.

—— PPC (Greece)

I'm Online Chat Now
Company News
FAQ for IP Water Ingress Protection Testing Equipment
Latest company news about FAQ for IP Water Ingress Protection Testing Equipment

Question 1: What should we do if the nozzles do not spray water?

Answer:

1). Firstly, user should check the water filter if it should be changed.

2).Secondly, the air compressor can be used to air dry the nozzles to check if they are clogged.

3). Replace with spare nozzles and try again.

 

Question 2: How to maintain the IPX34 osillating tubes?

Answer:

1). The waterproof test machine needs to be placed in a ventilated position as it has been dealing with water for year round.

2). A very important and easily ignored problem: remember to lock the wheels when placing the test equipment, so as to avoid the test equipment running around and affecting the accuracy of the test results.

3). IPX3-4 osillating tube is differ from IPX12 drip box, the nozzle is less than IPX12 drip box. Therefore, the equipment do not need air compressor for drying. Because of this, the durability of the IPX34 environmental rain test chamber is weakened.

4). Regularly to check the water filter at the back of the equipment, to avoid blocking the nozzles.

5). When the test equipment is often not used, the power should be unplugged to avoid long-term standby and increase the consumption of electronic components.

 

Question 3: How to operate the IPX9K high temperature and pressure jetproof testing equiment at the condition of 80℃ and 8000~10000Kpa?

Answer:

1). Power on the equipment and start the heating system.

2). Fix the specimen on the turntable

3). Adjust the distance between specimen and spray nozzle at the range of 100mm to 200mm ( the length of spray nozzle can be adjusted).

4). Wait for the water to heat up to 80℃, and select the test time for IPX9K test.

5). Wait for a period time to open the door after the test finished (The inner box will form a lot of water steam during the test which may cause scald).

6). Do not take out the specimen under tested by hand to avoid burns.

7). Take the specimen to lab for inpection.