Product Details:
Payment & Shipping Terms:
|
Standard: | UL94, IEC60695-11-4, IEC60695-11-3 | Control System: | PLC+touch Screen |
---|---|---|---|
Configuration: | Wrapping Tissue, Bunsen Burner, Pick-up Tray, Thermocouple | Working Power: | 220V/50Hz |
Warrranty: | 12 Months | ||
High Light: | flame test apparatus IEC60695-11-4/3,UL94 flame test apparatus,PLC Type flame detector tester |
PLC Type 50W and 500W Horizontal and Vertical Flame Tester According to UL94 and IEC60695-11-4/3
1. Introduction
Horizontal and vertical flame test apparatus is applied to 50W vertical flame test includes V-0, V-1, V-2, horizontal flame tester HB, and 500W vertical flame test includes 5VA, 5VB, HF-1, HF-2 and HBF. It is widely used in the R&D department, QC department and production department of lighting equipment, low-voltage apparatus, household appliance, motor and etc.. And also can be used in the industries of insulation material, plastic, non-metalic material, foam with a density of not less than 250kg/m' and other solid combustible material.
The horizontal and vertical flame test apparatus is controlled by 7-inch color touch screen and PLC. The test apparatus is built-in air exhaust, the front panel is transparent for easy observation of the experimental process, and the shell is made of painted coating metal sheet. Fully independent electrical circuit and gas circuit design for safer use.
Horizontal and vertical flame test apparatus is applied to simulate the fire at the early stage surround the electric and electronics appliance, and then evaluate the degree of fire hazard. It is designed according to UL94, IEC60695-11-4, IEC60695-11-3, IEC60950, IEC707, IEC60695-2-2, ISO1210 and etc.. The flammability test uses a bunsen burner of a specified size and specified gas source (methane or natural gas) to ignite the horizontal or vertical specimen for several times in a certain flame height and a certain flame angle, the fire hazard is evaluated by burning rate, after glow time, after flame time and damage length of the specimen.
2. Parameter
Control system | 7-inch color touch screen+PLC |
Burner | Φ9.5mm ± 0.3mm, 100mm length bunsen burner, source: 50W and 500W |
Dip angle | 00, 200, 450 (manual adjustment) |
Timer |
Integrated in PLC Accuracy: ±0.1s |
Number of flame | 0~9999 times (presettable) |
Fuel gas | 98% methane or 37MJ/m3±1MJ/M3 natural gas (prepare by user) |
Flow and pressure record | Double flowmeters and manometer |
Gas Pressure | 0.1Mpa |
Gas Flow | 50W:105±5ml/min or 18±1 mm, 500W:965±30ml/min or 117±3 mm |
Flame calibration system |
Complete the flame calibration with one key, the system automatically completes the flame calibration 3 times, and the system automatically judges each time. Any unqualified will prompt to re-adjust the flame, and stop the next flame calibration, if qualified, continue to the next flame calibration. 1). Flame temperature required: heat up from 1000C±50C to 7000C±30C within the time of 44s±2s (50W) and 54s±2s (500W) 2). Flame height: 50W: 20±2mm (50W) 500W: 40±2mm (inner), 125±10mm (outer) 3). Thermocouple: Φ0.5mm K type thermocouple (The insulating sleeve heat resist to 1250 ℃), 2 pieces to detect the temperature 4). Temperature measuring copper blocks |
Volume of chamber | 0.75 CBM, matte black background, buit-in exhaust fan, exhaust hole: Φ70mm |
Accessories | Pick-up tray, sample clamp, wrapping tissue 12g/m2 ~30g/m2, steel wire mesh, height gauge, cotton, remote-controller and etc. |
Dimension of apparatus | 1260mm(W)*700mm(D)*1300mm (H) |
Power supply | AC220V, 50Hz, ≤500W |
Contact Person: Ms. Penny Peng
Tel: +86-18979554054
Fax: 86--4008266163-29929