Product Details:
Payment & Shipping Terms:
|
Standard: | IEC60601-1 | Test Items: | AC Withstand Voltage, DC Withstand Voltage, Breakdown Current |
---|---|---|---|
Application: | Household Appliances, Medical Equipment, Lighting Industry | Test Voltage: | 0-5KV (AC/DC) |
Test Current Range: | 0.1-100mA | ||
Highlight: | 5KV AC DC Withstand Voltage Tester,Withstand Voltage Tester IEC60601-1,Medical Withstand Voltage Tester |
5KV AC/DC Withstand Voltage Tester for Testing Medical Equipment as IEC60601-1
1. Introduction
The RK2672CY medical withstand voltage tester has all the functions of the conventional withstand voltage tester, and has added an arc (flashover) detection function. Through an external oscilloscope, it can detect the “flashover” phenomenon intuitively, accurately, quickly and reliably. When the electrical equipment under test has no "flashover" phenomenon, the oscilloscope will display a stable "Lissajous figure (that is, a closed circle)", if the electrical equipment under test has a "flashover" phenomenon, the edge of the Lissajous waveform will appear jagged, indicating that there is a "flashover" inside the electrical equipment under test.
This product is also suitable for all kinds of electrical equipment that need to detect "flashover" defects. It is an indispensable withstand voltage tester for medical electrical equipment manufacturers, maintenance departments, users, product quality inspection departments, and technical supervision departments.
2. Application
1). Medical equipment: all kinds of new medical equipment and medical supporting equipment, heart monitoring, medical imaging, biochemical analysis equipment, sphygmomanometer and thermometer and other home medical equipment
2). Diagnosis and treatment equipment: X-ray diagnosis and inspection equipment, ultrasonic diagnosis, nuclear medicine, endoscope system, ENT medical equipment, dynamic analysis and medical equipment, cryogenic refrigeration equipment, dialysis medical equipment, first aid equipment
3). Hospital nursing equipment and appliances: all kinds of beds, cabinets, operating chairs, beds, etc.
4). Auxiliary equipment: medical nursing data and image processing equipment, rehabilitation equipment, special equipment for the disabled, etc.
5). Dental medical equipment: dental diagnostic medical equipment, dental surgical equipment, dental technical equipment, medical magnetic resonance equipment
3. Features
1). Test time up to 999s
2). Leakage current, voltage and time are simultaneously displayed
3). Leakage current, test time can be set continuously
4). With an oscilloscope interface, it can monitor the phenomenon of ignition, arcing and flashover of the measured object
5). GB9706.1-2020 (IEC60601-1: 2012) experimental voltage requirements: flashover or breakdown should not occur during the experiment. The oscilloscope interface is specially added to the medical withstand voltage tester, and the flashover and arcing phenomenon can be observed through the Lissajous figure
4. Parameters
Model | Rk2672CY | |
ACW | Output voltage range | (0.00-5.00)KV |
Maximum output power | 500VA (5.0KV, 100mA) | |
Maximum rated current | 100mA | |
Current gear | 2mA, 20mA, 100mA | |
Output waveform | Sine-wave | |
Output waveform distortion | ≤5% (no-load or pure resistive load) | |
Test time | 0.0s-999s 0=continuous testing | |
DCW | Output voltage range | (0.00-5.00)KV |
Maximum output power | 100VA (5.0KV, 20mA) | |
Maximum rated current | 20mA | |
Current gear | 2mA, 20mA | |
Ripple factor | ≤5% | |
Test time | 0.0s-999s 0=continuous testing | |
Voltmeter | Range | (0.00-5.00)KV |
Accuracy | ±(5%+3 digits) | |
Resolution | 10V | |
Display value | RMS | |
Ammeter | Range | Range 1: 0.1mA~2mA, range 2: 2mA~20mA, range 3: 20mA~100mA |
Accuracy | 2mA: 1μA, 20mA: 10mA; 100mA: 0.1mA | |
Resolution | ±(5%+3 digits) | |
Timer | Range | 0.0s-999s |
Resolution | 0.1St | |
Accuracy | ±(1%+50ms) | |
PLC interface | Optional | |
External interface | Yes | |
Dimension | D468mm*H196mm*W378mm | |
Weight | 20.65kg | |
Accessories | Power cable, HV test wire, ground wire. BNC wire, HV probe |