Product Details:
Payment & Shipping Terms:
|
Standard: | IEC60112 | Control System: | PLC |
---|---|---|---|
Configuration: | Platinum Electrodes, Dripping System, Glassware | Working Power: | 220V/50Hz |
Warrranty: | 12 Months | Application: | Flammability Testing |
Highlight: | IEC60112 Standard Tracking Index Test Device,PLC Control System Flammability Testing Device |
The tracking index test device is designed according to the standards IEC 60112, GB/T 4207, IEC62368-1 clause 5.4.3.3, IEC60598-1 clause 13.4, IEC60601-1 clause 8.9.1.7, IEC60335-1 clause 29, IEC61010-1 clause 6.7.1.3, IEC60884-1 clause 28.2, and BS1363-1 clause 8.2 Appendix C. It is widely used in material testing on lamps, household appliances, IT/AV appliances, and electrical accessories. The tracking resistance index measured by the tracking index tester can identify the level of tracking resistance of raw materials used in electrical equipment working in a humid environment. The tracking index test device is equipped with a 7-inch color touch screen, PLC, imported medical grade titration pump, high-power voltage regulator, and pure platinum electrodes.
- IEC60112 Method for the determination of the proof and the comparative tracking indices of solid insulating materials
- IEC60598-1 Luminaires – Part 1: General requirements and tests clause 13.4
- IEC62368-1 Audio/video, information and communication technology equipment – Part 1: Safety requirements clause 5.4.3.3
- IEC60601-1 Medical electrical equipment-Part 1: General requirements for basic safety and essential performance clause 8.9.1.7
- IEC60335-1 Household and similar electrical appliances – Safety –Part 1: General requirements clause 29
- IEC61010-1 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance clause 6.7.1.3
- IEC60884-1 Plugs and socket-outlets for household and similar purposes –Part 1: General requirements clause 28.2
- BS1363-1 13 A plugs, socket-outlets, adaptors and connection units Part 1: Specification for rewirable and non-rewirable 13 A fused plugs clause 8.2 Annex C
Control system | 7-inch color touch screen+PLC |
Electrode material | platinum (Pt) purity≥ 99% |
Electrode size | (2mm±0.1mm)×(5mm±0.1mm)×(40mm±5mm), length≥12mm, angle of chisel-edged: 30°±2° |
Electrode distance | 4.0mm±0.1mm, angle: 60°±5° |
Electrode pressure | 1.00±0.05N |
Electrode voltage | 100~600V |
Height of the drops | 35mm±5mm adjustable |
Size of drops | 20 drops: 0.380g~0.480g, 50 drops: 0.997g~1.147g |
Interval of the drops | 30s±5s, 50 drops: 24.5min±2min |
Test solutions resistance | Solution A: 0.1%NH4Cl, 3.95±0.05Ωm, solution B: 1.98±0.05Ωm |
Number of drops | 0~9999 (presettable, record and display) |
Test wind speed | 0.2m/s |
Short circuit current drop | 1.0A±0.1A drop 10% |
Leakage indicator | 0.50A±10% , 2.00s±10% |
Standard | IEC60112, UL746A, ASTM D 3638-92, DIN 53480 |
Accessories | Dripping system, glassware, block gauge, electrodes and etc. |
Outer dimension | 800mm (W)*500mm (D)*970mm (H) |
Working Power | 220V/50Hz, 1KVA |
Contact Person: Ms. Penny Peng
Tel: +86-18979554054
Fax: 86--4008266163-29929